CS-T240
Technical Spesification:
1 | Overall Performance | |
– Equipment Type | Fully Automatic, discrete, STAT priority | |
– Analysis Rate | Colorimetry speed 240T/H ( single/double reagent ) | |
– Test Principle | Colorimetry, turbidimetry | |
– Analysis Method | End-point, kinetics, fixed timer, etc | |
Support single/double wavelenght and 1-2 multiple reagent item, | ||
linear and non-linear calibration | ||
2 | Sample Reagent Unit | |
– Sample reagent position | The reagent and sample share one disk, totally 67 positions | |
continuously cooling at all positions to keep 5-150C within 24 hours | ||
– Sample cuvettes specification | Standard cup, original blood tube, multi specification tube available (ɸ 12-16)mm x (25-100)mm | |
– Sample Volume | 3µl – 35µl, 0,1µl stepping | |
– Reagent Volume | R1:10µl – 350µl, R2:10µl – 200µl, 1µl stepping | |
– Sample reagent probe | 1, with the function of liquid level detection and collision detection | |
– Sample reagent probe carring rate | Automatic warm water cleaning. Carring rate ≤ 0.1% | |
– Automatic sample dilution | 3-115 | |
3 | Reaction Unit | |
– Reaction cuvette | 120 positions optical plastic cup, optical diameter is 6 mm | |
– Total volume of reaction liquid | 150µl – 550µl | |
– Reaction temperature | 370C, ±0.10C | |
– Reaction disk constant temperature | Circulating water | |
– Mixer | 1, after joining reagent, blending immediately | |
– Reaction cuvette cleaning | 8 Stops 12 steps by warm water rinsing | |
– Wastewater treatment | With the function of concentrated waste liquid level alarming | |
4 | Optical System | |
– Light Source | 20W/12V halogen lamps | |
– Monochromator | Grating photometry | |
– Photoelectron road | After spectrophotometry | |
– Wavelengths | 340nm, 380nm, 405nm, 450nm, 480nm, 505nm, 546nm, | |
570nm, 600nm, 660nm, 700nm, 750nm or 800nm | ||
– Detector | Photodiode LED array | |
– OD linear range | 0 – 3.3abs | |
5 | Calibration and QC | |
– Calibration Method | 1 point linear method, 2 point linear method, multiple point linear method | |
non – linear method | ||
– Calibration tracking | Automatic description calibration K-value trends | |
– QC Method | Real time QC, indivudual QC and monthly QC | |
– Out of Control Processing | Alarming for out control sample, record lost control reason | |
6 | Operating System | |
– PC Operating system | Windows XP | |
– Analysis control software | Graphical operating software english version | |
– Main function of software | Automatic calibration, automatic barcode scanning, item compounding test, | |
reagent info management, serum index, whole reaction process monitoring, | ||
dirty cup memory evading, prevent cross-contamination procedure, patient information memory | ||
and association input, automatic report audit, data multiple parameter query, | ||
report format statistic and printing, reference range classification, alarming information | ||
classification, user operating right classification, automatic dormancy and wake, | ||
real-time online help | ||
– Report printing | Report formats support the user-defined mode, QC and state information etc. | |
– PC Configuration | CPU ≥2.2GHz(dual-core processor); Memory ≥1G; Harddisk ≥ 160G; | |
17 inch LCD display; Stylus, inkjet or laser printer (optional) | ||
– System connection | TCP/IP network connection, standard RS-232C | |
7 | Other | |
– Volume | 998mm x 752mm x 517mm | |
– Weight | 120Kg | |
– Power Supply | Voltage AC 220V ±22V, 50/60Hz, power 650VA |